Related Experiment Video
Updated: Jun 29, 2026

Wireless Telemetry Device Implantation in a Fontan Ovine Model for Continuous and Long-Term Hemodynamic Monitoring
Published on: May 2, 2025
Development of a fully implantable wireless pressure monitoring system
Robert Tan1, Timothy McClure, C K Lin
1Department of Bioengineering, University of California, Los Angeles, CA, 90095, USA.
Abstract:
A fully implantable wireless pressure sensor system was developed to monitor bladder pressures in vivo. The system comprises a small commercial pressure die connected via catheter to amplifying electronics, a microcontroller, wireless transmitter, battery, and a personal digital assistant (PDA) or computer to receive the wireless data. The sensor is fully implantable and transmits pressure data once every second with a pressure detection range of 1.5 psi gauge and a resolution of 0.02 psi. In vitro calibration measurements of the device showed a high degree of linearity and excellent temporal response. The implanted device performed continuously in vivo in several porcine studies lasting over 3 days. This system can be adapted for other pressure readings, as well as other vital sign measurements; it represents the first step in developing a ubiquitous sensing platform for telemedicine and remote patient monitoring.