Related Experiment Video
Updated: Jun 28, 2026

Uracil-DNA Glycosylase Assay by Matrix-assisted Laser Desorption/Ionization Time-of-flight Mass Spectrometry Analysis
Published on: April 22, 2022
Laboratory evaluation of the COBAS MIRA S random access analyser
T Kanluan1, S Intaramanee, S Tangvarasittichai
1Department of Clinical Chemistry Faculty of Medical Technology Mahidol University Bangkok 10700 Thailand.
Abstract:
The random access analyser COBAS MIRA S (Roche Diagnostics) was evaluated for two months. The instrument is a computer-controlled discrete analyser which can be run in a combination profile and single test mode. This instrument has special features, including an automatic cuvette segment changer, a reagent rack cooling system, an external keyboard and monitor, as well as a bar-code facility for the entry of test parameters, worklists and sample identification numbers. Study of within-run and between-run precision gave values of % CV 0.54-3.37 and 0.61-3.65, respectively, for a variety of assays. Linearity testing to the upper limit of each test was also studied and were found to cover the necessary pathological range. Within the two-month period, no major problems were encountered. The instrument required minimum operator attention during operation. Correlation studies with the Hitachi 705 using six clinical chemistry tests (glucose, cholesterol, triglyceride, ALP, AST, ALT) gave correlation coefficients ranging from 0.95-0.99 and slopes of 0.91-1.17.
