A low Z linac and flat panel imager: comparison with the conventional imaging approach
D A Roberts1, V N Hansen, A C Niven
1Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, Downs Road, Sutton, Surrey, UK.
Abstract:
Experimental and Monte Carlo simulations were conducted for an Elekta Ltd Precise Treatment System linac fitted with a low Z insert of sufficient thickness to remove all primary electrons. A variety of amorphous silicon based panels employing different scintillators were modelled to determine their response to a variety of x-ray spectra and produce an optimized portal imaging system. This study has shown that in a low Z configuration the vast majority of x-rays are produced in the nickel electron window, and with a combination of a carbon insert and caesium iodide based XVI-panel, significant improvement in the object contrast was achieved. For thin, head and neck-type geometries, contrast is 4.62 times greater for 1.6 cm bone in 5.8 cm water than the standard 6 MV/iViewGT system. For thicker, pelvis-type geometries contrast increases by a factor of 1.3 for 1.6 cm of bone in 25.8 cm water. To obtain images with the same signal-to-noise ratio as the 6 MV/iViewGT system, dose reductions of a factor of 15 and 4.2 are possible for 5.8 cm and 25.8 cm phantoms respectively. This design has the advantage of being easily implemented on a standard linac and provides a portal image directly from the therapy beam aperture.
Related Concept Videos
Confocal Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...