Compression feedback devices over estimate chest compression depth when performed on a bed
Related Experiment Video
Updated: Jun 28, 2026

Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Gavin D Perkins1, Laura Kocierz, Samuel C L Smith
1Warwick Medical School, University of Warwick, Warwick CV4 7AL, United Kingdom. g.d.perkins@warwick.ac.uk
Introduction:
CPR feedback/prompt devices are being used increasingly to guide CPR performance in clinical practice. A potential limitation of these devices is that they may fail to measure the amount of mattress compression when CPR is performed on a bed. The aim of this study is to quantify the amount of mattress compression compared to chest compression using a commercially available compression sensor (Q-CPR, Laerdal, UK). A secondary aim was to evaluate if placing a backboard beneath the victim would alter the degree of mattress compression.
Methods:
CPR was performed on a manikin on the floor and on a bed with a foam or inflatable mattress with and without a backboard. Chest and mattress compression depths were measured by an accelerometer placed on the manikin's chest (total compression depth) and sternal-spinal (chest) compression by manikin sensors.
Results:
Feedback provided by the accelerometer device led to significant under compression of the chest when CPR was performed on a bed with a foam 26.2 (2.2)mm or inflatable mattress 32.2 (1.16)mm. The use of a narrow backboard increased chest compression depth by 1.9mm (95% CI 0.1-3.7mm; P=0.03) and wide backboard by 2.6mm (95% CI 0.9-4.5mm; P=0.013). Under compression occurred as the device failed to compensate for compression of the underlying mattress, which represented 35-40% of total compression depth.
Conclusion:
The use of CPR feedback devices that do not correct for compression of an underlying mattress may lead to significant under compression of the chest during CPR.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
