[Falsely elevated-bispectral index]
A Odri1, A Cavalcanti, V Billard
1Département d'anesthésie, institut Gustave-Roussy, 39, rue Camille-Desmoulins, 94805 Villejuif cedex, France. blanieodri@wanadoo.fr
Abstract:
We report a case of a falsely elevated-bispectral index (BIS) during a general anaesthesia combining remifentanil TCI, desflurane and nitrous oxide for an isolated-limb chemotherapy. During surgery, BIS increased and stabilized around 70, with neither residual neuromuscular blockade nor clinical sign of awareness. These high BIS values were attributed to high-electromyographic activity and electric artefacts, such as extracorporeal-circulation machine and tourniquet. At the end of the surgery, the BIS returned to expected values around 50. The patient did not complain of intraoperative recall. This case reminds us that the BIS has some limits as being sensitive to EMG or environment artefacts that should be eliminated before deepening anesthesia. To do so, a decision algorithm is proposed that may be used for all situations of surprising high BIS, taking into account the level of neuromuscular blockade, clinical response to orders and the presence of devices likely to induce electrical or mechanical artefacts.
More Related Videos
10:21Evaluation of Hydration Status by Bioelectrical Impedance Vector Analysis in Patients with Ischemic Heart Disease Undergoing Exercise Stress Test
Published on: September 22, 2023
04:05Clinical Application of Phase Angle and BIVA Z-Score Analyses in Patients Admitted to an Emergency Department with Acute Heart Failure
Published on: June 30, 2023
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Measurement of Blood Pressure