Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Use of an electrostatic preconcentration system to identify interferences in atomic-absorption spectrometry
1Trace Analysis Research Centre, Department of Chemistry, Dalhousie University, Halifax, N.S., Canada.
Abstract:
Interference effects can be identified by deviations from ideality of the slopes of the concentration curves generated with an electrostatic trapping system. The technique is employed to detect spectral interferences in atomic-absorption measurements, due to the presence of sodium chloride, and caused by imperfect background correction. The electrostatic trap is found to be a particularly convenient means of obtaining the required concentration curves because the device gives a ready control of the concentration factor, reliable performance for a range of different solutions, and ease of use in dealing with the necessary number of samples.
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Atomic Emission Spectroscopy: Interference
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Lab
Atomic Emission Spectroscopy: Overview