Single-point titration of metal ions and ligands by measuring change in pH
A Lewenstam1, A Ivaska, E Wänninen
1Department of Analytical Chemistry, Abo Akademi, 20500 Abo Turku, Finland.
Abstract:
A single-point titration method is described for the determination of metal ions and ligands. The method is based on deprotonation of a ligand on complexation with a metal. The resultant change in pH is used to calculate the concentration of the analyte. The limits and advantages of the method are discussed. Four selected systems, Pb(2+)-HPO(2-)(4), Cu(2+)-HCO(-)(3), Cd(2+)-EDTA and Ni(2+)-ethylenediamine are used to demonstrate the validity of the equations. The method is applied to determination of total carbonate in serum.
More Related Videos
Related Concept Videos
Complexometric Titration: Overview
Complexometric Titration: Ligands
EDTA: Indirect and Alkalimetric Titration
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides and...
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...