Related Experiment Video
Updated: Jun 28, 2026

Post Column Derivatization Using Reaction Flow High Performance Liquid Chromatography Columns
Published on: April 26, 2016
Determination of the flotation collector ethyl xanthate by flow-injection analysis
M Kopanica1, V Stará, A Trojánek
1UNESCO Laboratory of Environmental Electrochemistry, J. Heyrovský Institute of Physical Chemistry and Electrochemistry, Czechoslovak Academy of Sciences, Dolejskova 3, 18223 Prague 8, Czechoslovakia.
Abstract:
A flow-injection method for determination of xanthate with amperometric detection is presented. A composite carbon electrode modified with silica gel is used for detection. Very good reproducibility (1.5% r.s.d.) results from renovation of the electrode surface by three potential scans between measurements. In the analysis of flotation liquor, the interfering effect of metal complexes, formed by leaching of the ore in water, is eliminated by addition of EDTA. The method can be used to determine ethyl xanthate in the concentration range 0.10-10.0 mg/1, with a detection limit of 0.04 mg/1.