Related Experiment Video
Updated: Jun 28, 2026

Separation of Uranium and Thorium for 230Th-U Dating of Submarine Hydrothermal Sulfides
Published on: May 20, 2019
Fluorimetric determination of uranium(VI) in waters by flow-injection analysis after preconcentration on a silica gel
J Havel1, M Vrchlabský, Z Kohn
1Department of Analytical Chemistry, Faculty of Science, Masaryk University, Kotlárská 2, 611 37 Brno, Czechoslovakia.
Abstract:
A method was developed for the selective determination of trace concentrations of uranium(VI) by flow-injection analysis (FIA) with fluorimetric detection. Uranium(VI) is selectively separated and/or pre-concentrated from a volume up to 20 ml on an activated silica gel microcolumn (2 x 40 mm) from a medium of 0.03M EDTA, 0.06M tartrate, and/or 0.05M NaF at pH = 9.3. After washing the column the uranium is eluted with a mixture of 1.33M sulphuric and phosphoric acids and determined with a relative standard deviation not exceeding 6% for concentrations in the range 10-250 mug/l. The detection limit was estimated to be 0.1-0.2 mug of uranium. The method has been verified on artificial water samples with high content of the interfering elements and applied to analysis of waste and natural waters.
Related Concept Videos
Microbial Bioremediation of Uranium
Atomic Fluorescence Spectroscopy