Related Experiment Video
Updated: Jun 28, 2026

Light Enhanced Hydrofluoric Acid Passivation: A Sensitive Technique for Detecting Bulk Silicon Defects
Published on: January 4, 2016
Silicon determination by inductively coupled plasma atomic emission spectrometry after generation of volatile silicon
A Lopez Molinero1, L Martinez, A Villareal
1Department of Analytical Chemistry, Faculty of Sciences, University of Zaragoza, 50009 Zaragoza, Spain.
Abstract:
A method for the determination of silicon by inductively coupled plasma atomic emission spectrometry (ICP-AES) is described. The procedure is based on a discontinuous generation of volatile silicon tetrafluoride in concentrated sulphuric acid medium after injecting 125 mul of 0.1%, w/v sodium fluoride solution into 100 mul of the sample. The gaseous silicon tetrafluoride is fed directly into the ICP torch by a flow of 250 ml min(-1) Ar carrier gas. The calibration curve was linear up to at least 100 mug ml(-1) of Si(IV) and the absolute detection limit was 9.8 ng working with a solution volume of 100 mul. The relative standard deviation for six measurements of 10 mug ml(-1) of Si(IV) was 2.32%. The method was applied to the determination of silicon in water and iron ores.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Emission Spectroscopy: Lab
Types of Semiconductors
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.