T O Hirche1, T Born, S Jungblut
1Department of Pulmonary Medicine, University Hospital Frankfurt, Theodor Stern Kai 7, 60590 Frankfurt, Germany.
Background:
Oxy-Gen lite, a recently developed combined electrolysis and fuel cell technology, de-novo generates oxygen with high purity for medical use from distilled water and room air. However, its use in patients with chronic respiratory failure has never been evaluated.
Objectives:
To test the clinical applicability and safety of Oxy-Gen lite technology, we enrolled 32 COPD patients with chronic hypoxemia and long-term oxygen therapy (LTOT) in a controlled, randomized, multicenter clinical trial.
Materials And Methods:
Standard continuous oxygen therapy with a maximal flow rate of 2 L/min was tested against pulsatile oxygen delivery by Oxy-Gen lite. Oxygen saturation at seated-rest was recorded over 30 min and used as a primary read-out parameter. Oxygen saturation was also recorded during mild physical strain (speaking out loud) or overnight's sleep.
Results:
Both methods of oxygen supply established oxygen saturations within the normal range (i.e., upper plateau of the sigmoid oxyhaemoglobin dissociation curve) compared to breathing room air (p<0.0001). Mean oxygen saturation under standard continuous oxygen flow or Oxy-Gen lite technology during rest, physical strain or sleep proved statistically equivalent (95%CI<2.5% of reference saturation).
Conclusion:
The use of Oxy-Gen lite in COPD patients with hypoxemia and LTOT
Related Concept Videos
Treatment for Pulmonary Arterial Hypertension: Oxygen Therapy for Respiratory Failure
Oxygen therapy is vital in increasing and maintaining blood oxygen levels in PAH patients. As a result, it aids in reducing fatigue, improving...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Microbial Fuel Cells
Batteries and Fuel Cells
Oxygenic Photosynthesis
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...