Microbiologic contamination of automatic injectors at MDCT: experimental and clinical investigations

Boris Buerke1, Alexander Mellmann, Christoph Stehling

  • 1Department of Clinical Radiology, University of Muenster, Albert-Schweitzer-Strasse 33, 48149 Muenster, Germany. buerkeb@uni-muenster.de

Abstract

Insights

Multiple uses of contrast agent injector syringes in MDCT can lead to bacterial contamination, posing a risk of septic complications. To ensure patient safety, single-use syringes are recommended, even with optimized hygiene protocols.

Area of Science:

  • Medical Imaging
  • Infection Control
  • Radiology

Background:

  • Antiseptic skills are crucial for staff using contrast agent injectors.
  • Microbiologic contamination of injector syringes is a potential risk in MDCT procedures.

Purpose of the Study:

  • To evaluate the risk of microbiologic contamination of contrast agent and saline solution injector syringes.
  • To assess contamination in both experimental and clinical settings during MDCT procedures.

Main Methods:

  • Simulated clinical setting to investigate contamination of CT injector syringes over multiple uses.
  • Microbiologic analysis of devices and staff palms via imprints during clinical routine.
  • Evaluation of hygienic conditions and staff training on antiseptic behaviors.

Main Results:

  • Experimental simulations showed initial sterility followed by contamination with dermal bacteria (staphylococci).
  • Clinical settings revealed contamination with skin and oral flora on devices and staff palms.
  • No nosocomial or fecal pathogens were detected on surfaces or palms.

Conclusions:

  • Environmental optimization alone does not prevent syringe contamination.
  • Prohibiting multiple uses of syringes for different patients is essential.
  • Single-patient use of syringes minimizes the risk of septic complications.