An improved method to validate the relative humidity generation in sorption balances
Lars Wadsö1, Anders Anderberg, Ingrid Slund
1Building Materials, Lund University, Lund, Sweden. lars.wadso@byggtek.lth.se
Summary
This study presents a new method for validating relative humidity (RH) generation in sorption balance instruments. The improved approach offers a faster and more reliable way to determine the deliquescence point of salts.
Area of Science:
- Materials Science
- Physical Chemistry
- Analytical Chemistry
Background:
- Sorption balances are crucial for studying solid-vapor interactions and measuring sorption isotherms.
- Accurate relative humidity (RH) generation is essential for the reliable performance of sorption balance instruments.
- Current validation methods often involve slow RH ramps or steps past salt deliquescence points.
Purpose of the Study:
- To introduce an improved method for validating the relative humidity (RH) generation in sorption balance instruments.
- To demonstrate an alternative approach for determining the deliquescence relative humidity (RH) of salts.
- To offer a less time-consuming and more disturbance-resilient validation technique.
Main Methods:
- Utilizing sorption balances to weigh samples exposed to programmed relative humidity (RH).
- Implementing a stepwise constant RH approach at quasi-random RH values above and below the deliquescence RH.
- Analyzing the rates of mass change as a function of RH to determine the deliquescence point.
Main Results:
- The improved method accurately determines the deliquescence RH of salts.
- Results obtained are comparable to traditional slow ramp methods.
- The new approach exhibits reduced sensitivity to environmental disturbances.
Conclusions:
- The presented stepwise constant RH method provides a valid and improved alternative for sorption balance performance validation.
- This technique is more robust and efficient than existing methods.
- Accurate RH control is critical for reliable solid-vapor interaction studies.
Related Concept Videos
Volatilization
Volatilization gravimetry is an analytical technique that measures the mass lost due to the volatilization of the substance. This technique is used to estimate the amount of volatile material in a sample. To perform this method, heat a known amount of the sample to a high temperature in a crucible or other suitable vessel. The volatile substance in the sample evaporates, and the vapor is completely expelled from the crucible either by heating the sample or bubbling a stream of inert gas through...
Precipitation Titration: Endpoint Detection Methods
In argentometric precipitation titrations, endpoints can be detected visually by the Mohr, Volhard, and Fajans methods. In the Mohr method, adding a soluble chromate indicator gives an initial yellow color to the analyte solution. As the titrant is added, the first excess of silver ions forms a red silver chromate precipitate, marking the endpoint. The solution pH should be maintained at about 8 by adding solid CaCO3.
In the Volhard method, a standard excess of AgNO3 is first added to the...
In the Volhard method, a standard excess of AgNO3 is first added to the...
Measurement of Air Content in Concrete
Air content measurement in concrete is critical for ensuring structural integrity and durability of concrete structures, especially in environments prone to severe weather conditions. Accurate air content analysis optimizes concrete's resistance to freeze-thaw cycles and enhances its workability and strength. Several methods are standardized under ASTM guidelines to measure the air content in fresh concrete, each suitable for different concrete types and conditions.
The pressure method,...
The pressure method,...