Optimal echo spacing for multi-echo imaging measurements of Bi-exponential T2 relaxation
Adrienne N Dula1, Daniel F Gochberg, Mark D Does
1Department of Biomedical Engineering, Vanderbilt University, Nashville, TN 37232-2310, United States.
Abstract:
Calculations, analytical solutions, and simulations were used to investigate the trade-off of echo spacing and receiver bandwidth for the characterization of bi-exponential transverse relaxation using a multi-echo imaging pulse sequence. The Cramer-Rao lower bound of the standard deviation of the four parameters of a two-pool model was computed for a wide range of component T(2) values and echo spacing. The results demonstrate that optimal echo spacing (TE(opt)) is not generally the minimal available given other pulse sequence constraints. The TE(opt) increases with increasing value of the short T(2) time constant and decreases as the ratio of the long and short time constant decreases. A simple model of TE(opt) as a function of the two T(2) time constants and four empirically derived scalars is presented.