Spatial resolution study and power calibration of the high-k scattering system on NSTX
1Department of Physics, POSTECH, Pohang 790-784, Republic of Korea. woochang@pppl.gov
The Review of Scientific Instruments
|December 3, 2008
Abstract:
NSTX high-k scattering system has been extensively utilized in studying the microturbulence and coherent waves. An absolute calibration of the scattering system was performed employing a new millimeter-wave source and calibrated attenuators. One of the key parameters essential for the calibration of the multichannel scattering system is the interaction length. This interaction length is significantly different from the conventional one due to the curvature and magnetic shear effect.