Characterization of cross-section correction to charge exchange recombination spectroscopy rotation measurements
W M Solomon1, K H Burrell, R Feder
1Princeton Plasma Physics Laboratory, Princeton University, Princeton, New Jersey 08543, USA. wsolomon@pppl.gov
Abstract:
Measurements of rotation using charge exchange recombination spectroscopy can be affected by the energy dependence of the charge exchange cross section. On DIII-D, the associated correction to the rotation can exceed 100 kms at high temperatures. In reactor-relevant low rotation conditions, the correction can be several times larger than the actual plasma rotation and therefore must be carefully validated. New chords have been added to the DIII-D CER diagnostic to view the counter-neutral-beam line. The addition of these views allows determination of the toroidal rotation without depending on detailed atomic physics calculations, while also allowing experimental characterization of the atomic physics. A database of rotation comparisons from the two views shows that the calculated cross-section correction can adequately describe the measurements, although there is a tendency for "overcorrection." In cases where accuracy better than about 15% is desired, relying on calculation of the cross-section correction may be insufficient.
More Related Videos
09:00Experimental Methods for Spin- and Angle-Resolved Photoemission Spectroscopy Combined with Polarization-Variable Laser
Published on: June 28, 2018
07:24Quantitative Atomic-Site Analysis of Functional Dopants/Point Defects in Crystalline Materials by Electron-Channeling-Enhanced Microanalysis
Published on: May 10, 2021
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
NMR Spectroscopy: Spin–Spin Coupling
π Electron Effects on Chemical Shift: Overview
Atomic Emission Spectroscopy: Interference
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Atomic Emission Spectroscopy: Instrumentation