Absolute calibration of the Lyman-α measurement apparatus at DIII-D
F M Laggner1, A Bortolon1, A M Rosenthal2
1Princeton Plasma Physics Laboratory, Princeton, New Jersey 08543, USA.
Abstract:
The LLAMA (Lyman-Alpha Measurement Apparatus) diagnostic was recently installed on the DIII-D tokamak [Rosenthal et al., Rev. Sci. Instrum. (submitted) (2020)]. LLAMA is a pinhole camera system with a narrow band Bragg mirror, a bandpass interference filter, and an absolute extreme ultraviolet photodiode detector array, which measures the Ly-α brightness in the toroidal direction on the inboard, high field side (HFS) and outboard, low field side (LFS). This contribution presents a setup and a procedure for an absolute calibration near the Ly-α line at 121.6 nm. The LLAMA in-vacuum components are designed as a compact, transferable setup that can be mounted in an ex situ vacuum enclosure that is equipped with an absolutely calibrated Ly-α source. The spectral purity and stability of the Ly-α source are characterized using a vacuum ultraviolet spectrometer, while the Ly-α source brightness is measured by a NIST-calibrated photodiode. The non-uniform nature of the Ly-α source emission was overcome by performing a calibration procedure that scans the Ly-α source position and employs a numerical optimization to determine the emission pattern. Nominal and measured calibration factors are determined and compared, showing agreement within their uncertainties. A first conversion of the measured signal obtained from DIII-D indicates that the Ly-α brightness on the HFS and LFS is on the order of 1020 Ph sr-1 m-2 s-1. The established calibration setup and procedure will be regularly used to re-calibrate the LLAMA during DIII-D vents to monitor possible degradation of optical components and detectors.
More Related Videos
09:40Measurement and Analysis of Atomic Hydrogen and Diatomic Molecular AlO, C2, CN, and TiO Spectra Following Laser-induced Optical Breakdown
Published on: February 14, 2014
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Uncertainty in Measurement: Reading Instruments
Accuracy and Precision
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Differential Leveling