Influence of a fluoridated medium with different pHs on commercially pure titanium-based implants
Rafael Sartori1, Cassia Bellotto Correa, Elcio Marcantonio
1Department of Periodontology, School of Dentistry at Araraquara, UNESP, Sao Paulo State University, Sao Paulo, Brazil.
Summary
Fluoride mouth rinses at different pH levels do not affect the corrosion resistance of titanium dental implants. This study found no evidence of corrosion, indicating implants maintain their integrity with these products.
Area of Science:
- Biomaterials Science
- Dental Materials
- Corrosion Engineering
Background:
- Titanium (Ti) dental implants are widely used due to their biocompatibility and mechanical properties.
- The long-term stability of dental implants in the oral environment is crucial for successful osseointegration.
- Exposure to oral hygiene products, such as fluoride mouth rinses, is a common factor influencing implant surfaces.
Purpose of the Study:
- To evaluate the effect of fluoride mouth rinses with varying pH levels on the corrosion resistance of commercially pure titanium (cp-Ti) dental implants.
- To investigate potential surface changes on titanium implants when exposed to simulated long-term use of fluoride rinses.
- To determine if fluoride concentration and pH influence the corrosion behavior of different titanium implant brands.
Main Methods:
- Forty-two cp-Ti dental implants from three commercial brands were subjected to simulated use.
- Specimens were immersed in a sodium fluoride (NaF) 1500 ppm solution at two different pH levels for 184 hours.
- Surface analysis was performed using Scanning Electron Microscopy (SEM) and Energy Dispersive Spectroscopy (EDS).
Main Results:
- SEM and EDS analyses revealed no signs of corrosion on the titanium implant surfaces.
- No evidence of fluoride ion incorporation into the implant surface was detected after exposure.
- The tested fluoride concentrations and pH values did not induce any measurable corrosion.
Conclusions:
- The pH of fluoride solutions and fluoride concentration do not impact the corrosion resistance of titanium dental implants.
- Commercially pure titanium dental implants demonstrate excellent resistance to corrosion from common fluoride mouth rinses.
- Clinical use of fluoride mouth rinses at the tested concentrations and pH is unlikely to compromise the structural integrity of titanium implants.