Evaluating the accuracy of an aneroid sphygmomanometer in a clinical trial setting
Yong Ma1, Marinella Temprosa, Sarah Fowler
1The George Washington University, Biostatistics Center, Rockville, Maryland, USA. dppmail@biostat.bsc.gwu.edu
Background:
The mercury sphygmomanometer, the "gold" standard for blood pressure measurements, has been gradually phased out in many institutions because of environmental concerns. Our on-going clinical trial compared the use of mercury vs. aneroid sphygmomanometers, before implementing a study-wide transition to the aneroid sphygmomanometer.
Methods:
The Diabetes Prevention Program Outcomes Study (DPPOS) studied the accuracy of the Welch Allyn Tycos 767-Series Mobile aneroid sphygmomanometer from 20 March 2006 to 21 June 2006 at multiple clinic centers. We compared readings from 997 participants in 24 clinic centers using both mercury and aneroid sphygmomanometers.
Results:
The study found no statistically significant difference for systolic blood pressure (SBP) (P > 0.05) and a small but significantly (P < 0.0001) lower (0.8 mm Hg) reading for diastolic blood pressure (DBP) using the aneroid sphygmomanometer. Regression analysis of aneroid vs. mercury showed regression lines (Y = 4.8 + 0.96X for SBP, Y = 3.1 + 0.95X for DBP) slightly but statistically significantly different from the line of equality (P < 0.001). Participants' age, sex, race/ethnicity, body mass index, blood pressure, and clinical center together explain about 8-10% of the variation of the difference between readings from the two sphygmomanometers. Based on the above result, on 1 August 2006, DPPOS clinics began the conversion from mercury to aneroid sphygmomanometers.
Conclusions:
The Welch Allyn Tycos 767-Series Mobile Aneroid model 7670-04 tested in this validation study can be used to replace mercury model in clinical trials.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...