Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
Light-scattering characteristics of explanted opacified Aquasense intraocular lenses
I J E van der Meulen1, H Porooshani, T J T P van den Berg
1Department of Ophthalmology, Academic Medical Center, Amsterdam, The Netherlands. i.j.vandermeulen@amc.uva.nl
Aim:
To study forward light-scattering characteristics of calcified explanted intraocular lenses (IOLs) (Aquasense, Ophthalmic Innovation International Ontario).
Methods:
The amount of light scattered by the opacified IOLs was measured using a validated in vitro set-up for angles from 1.7 degrees to 22 degrees . This set-up gives results directly comparable with straylight values as valid for the in vivo situation.
Results:
Straylight is highest at large angles and declines steeply approaching 0 degrees angle. This corresponds to the in vivo findings that opacified IOLs cause important visual complaints but have little effect on visual acuity. At 7.5 degrees , log (s) is around 1.8 and 2.9 for the two lenses respectively. This corresponds to 8 x and 100 x increases in straylight values compared with values in young, normal eyes.
Conclusion:
High straylight values caused by opacified IOLs can explain subjective complaints of reduced quality of vision in patients with opacified implants, despite good visual acuity.