[Reasons and complications of pacemaker replacement operation: clinical analysis of 69 case-times]
Xue-ling Liu1, Li-hui Ren, Hui-ming Ye
1Department of Cardiology, Shijitan Hospital, Beijing 100038, China.
Objective:
To analyze the reasons of cardiac permanent pacemaker replacement and the strategies to prevent relevant complications.
Methods:
The clinical data of 57 patients, 38 males and 19 females, aged 74 +/- 8 (56-94), 31 with sick sinus syndrome, 26 with II or III degree atrioventricular block, who underwent 63 times of cardiac permanent pacemaker replacement, including 13 times of lead replacement, were analyzed.
Results:
The reason of replacement included battery normal exhaustion for 57 case-times, battery exhaustion before the defined schedule for 2 case-times, lead electrodes incompletely fractured for 2 case-times, and infection of pacemaker pocket for 2 case-times. There were 9 cases of placement-related complications: pacemaker pocket hematoma (n=4), lead dislocation (n=3), and abandoned leads falling into the right ventricle (n=2). The average lifetime of old pacemakers was 9. 25 years (2 -15 years). The pacing threshold of ventricular leads after pacemaker replacement was (0.77 +/- 0. 40)V, significantly higher than the initial pacing threshold [(0.60 +/- 0.21)V, P < 0.01]. There were no significant differences in the lead impedance and R wave amplitude between the pacemaker replacement and initial implantation [(854 +/- 136)omega vs. (828 +/- 176)omega, and (12 +/- 4)mV vs. (12 +/- 4)mV, both P > 0.05].
Conclusion:
The main reason of pacemaker replacement is battery exhaustion. Most implanted ventricular leads still can be used. A rare serious complication of cardiac pacemaker replacement operation is abandoned lead falling into the right ventricle, and correct disposing of initial leads help avoid this complication .
More Related Videos
08:56Sterile Pericarditis in Aachener Minipigs As a Model for Atrial Myopathy and Atrial Fibrillation
Published on: September 24, 2021
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiomyopathy V: Interprofessional Care
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Cardiomyopathy III: Hypertrophic Cardiomyopathy