Harmonic long shears further reduce operation time in transanal endoscopic microsurgery
Pleun E A Hermsen1, Ifesegun D Ayodeji, Wim H C Hop
1Department of Surgery, IJsselland Hospital, Prins Constantijnweg 2, Capelle aan den IJssel, 2906 ZC, The Netherlands.
Surgical Endoscopy
|December 11, 2008
Summary
The new 5-mm harmonic long shears (H-LS) significantly reduce operation time and blood loss during transanal endoscopic microsurgery (TEM) compared to regular harmonic shears (HS). H-LS also enable complete resection in most cases, minimizing the need for additional instruments.
Area of Science:
- Colorectal Surgery
- Minimally Invasive Procedures
- Surgical Instrumentation
Background:
- Transanal endoscopic microsurgery (TEM) often requires additional instruments when using standard 5-mm harmonic shears (HS).
- Previous research suggests HS reduce operation time compared to diathermia in TEM.
- The efficacy of new 5-mm harmonic long shears (H-LS) for TEM has not been fully evaluated.
Purpose of the Study:
- To compare the effectiveness of 5-mm harmonic long shears (H-LS) versus regular harmonic shears (HS) in transanal endoscopic microsurgery (TEM).
- To assess the impact of H-LS on operation time, completeness of resection, and blood loss during TEM.
Main Methods:
- A retrospective study comparing 162 tumor excisions in 161 patients undergoing TEM between 2001 and 2006.
- Eighty-eight resections were performed using HS, and 74 using H-LS.
- Patient and tumor characteristics were analyzed, with a focus on operation time, blood loss, and resection completeness.
Main Results:
- Tumors resected with H-LS were smaller on average (34.4 cm²) than those with HS (44.1 cm²).
- Mean operation time was significantly reduced with H-LS (41 min) compared to HS (54 min), a 14% reduction after area correction.
- H-LS achieved complete resection in 88% of cases, versus 26% for HS. Blood loss was also significantly lower with H-LS (3 cc vs. 16 cc).
Conclusions:
- 5-mm harmonic long shears (H-LS) are more effective than regular harmonic shears (HS) for transanal endoscopic microsurgery (TEM).
- H-LS reduce operation time and blood loss, and facilitate complete tumor resection, often without requiring additional instruments.
- The findings support the use of H-LS for improved efficiency and outcomes in TEM procedures.