Related Experiment Video
Updated: Jun 27, 2026

Precise Electrochemical Sizing of Individual Electro-Inactive Particles
Published on: August 4, 2023
Voltammetric characterisation of silicon-based microelectrode arrays and their application to mercury-free stripping
Alfonso Berduque1, Yvonne H Lanyon, Valerio Beni
1Tyndall National Institute, Lee Maltings, University College, Cork, Ireland.
Abstract:
This paper describes the electrochemical characterisation of a range of gold and platinum microelectrode arrays (MEAs) fabricated by standard photolithographic methods. The inter-electrode spacing, geometry, numbers and dimensions of the electrodes in the arrays were found to influence the voltammetric behaviours obtained. Excellent correlation was found between experimental data and theoretical predictions employing published models of microelectrode behaviour. Gold MEAs were evaluated for their applicability to copper determination in a soil extract sample, where agreement was found between the standard analytical method and a method based on underpotential deposition-anodic stripping voltammetry (UPD-ASV) at the MEAs, offering a mercury-free alternative for copper sensing.
Related Concept Videos
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetric Techniques: Linear-Scan (E vs Time)
Voltammetric Techniques: Cyclic Voltammetry
Voltammetric Techniques: Pulse Voltammetry
Electrodeposition
Electrodeposition can...