Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
Reintroduction of Riva-Rocci measurements to determine systolic blood pressure?
E Verrij1, G van Montfrans, W J Bos
1Department of Internal Medicine, St Antonius Hospital, Utrecht, the Netherlands.
Introduction:
In 1896, Riva-Rocci introduced the upper arm cuff to measure systolic blood pressure. In 1905, Nicolai Sergeivich Korotkoff added the auscultatory technique, allowing measurement of both systolic and diastolic blood pressure. Both methods have, to our knowledge, never been formally tested against each other. In this study, we want to fill this gap in history.
Methods:
We measured systolic blood pressure by the Korotkoff sound technique and approximated the Riva-Rocci technique by measuring cuff pressure at the moment that the first pulsation became visible in Finapres readings, at the finger. This proxy of the Riva-Rocci technique allows an objective, offline, analysis. Measurements were performed simultaneously on the same arm in 57 subjects.
Results:
Systolic blood pressure measured by the Korotkoff sound technique was 167+/-30 mmHg (mean +/-SD ). Systolic blood pressure according to the Riva-Rocci technique was 165+/-32 mmHg. The Riva-Rocci technique underestimated measurements with the Korotkoff technique by 1.8+/-4.4 mmHg (NS, p=0.79).
Conclusion:
Riva-Rocci measurements of systolic blood pressure may be as good as the traditionally used Korotkoff measurements.
More Related Videos
Related Concept Videos
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: