Related Experiment Video
Updated: Jun 26, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
A tapered-sleeve transcortical pin external skeletal fixation device for use in horses: development, application, and
David M Nunamaker1, Ronald A Nash
1Richard S. Reynolds, Jr. Comparative Orthopedic Research Laboratory, Department of Clinical Studies, New Bolton Center, School of Veterinary Medicine, University of Pennsylvania, Philadelphia, PA 19348, USA. dmn@vet.upenn.edu
Objective:
To report development, configuration, application, and results of a tapered-sleeve transcortical pin external skeletal fixation device (TSP ESFD) for use on fractures of the distal aspect of the equine limb.
Study Design:
Optimization analysis of a TSP ESFD was carried out with mechanical testing and review of clinical case outcomes.
Animals:
Cadaveric adult third metacarpal bones (MC3) for mechanical testing; horses (n=7) with severely comminuted proximal (6; P1) or second (1) phalanx fractures.
Methods:
Mechanical testing of methods for attachment of TSP to the sidebars were tested as well as optimization of pin diameters. Outcome of clinical cases managed with the TSP ESFD were compared with outcomes of horses treated with previous ESFD.
Results:
A TSP ESFD using 7.94 mm diameter pins was used. Survival rate was 71%. One horse with an infected P1 fracture and contralateral laminitis subsequently fractured MC3 through the distal pin hole while wearing the TSP ESFD. One mare with contralateral laminitis was euthanatized. Three horses had open fractures.
Conclusions:
Design improvements incorporated into the TSP ESFD should increase treatment success.
Clinical Relevance:
External skeletal fixation continues to be a viable treatment alternative for severe injuries of the distal aspect of the equine limb. TSP ESFD should improve on the success of treatment by decreasing complications of bone fracture at the pin interface.