Related Experiment Video
Updated: Jun 26, 2026

09:31
Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Early failure of a nonmodular titanium femoral stem after primary hip arthroplasty
Khalid A Azzam1, Matthew S Austin, Peter F Sharkey
1Rothman Institute, Thomas Jefferson University, Philadelphia, PA 19107, USA.
The Journal of Arthroplasty
|January 20, 2009
Abstract:
Fully porous-coated cementless implants have a low failure rate and excellent survivorship in primary and revision surgery. Failure secondary to implant fracture is extremely rare with modern implant designs. We report an early fracture of a fully porous-coated femoral stem after primary total hip arthroplasty. The implant failed because of a filling mismatch between the proximal and distal femur.