Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
Nexfin noninvasive continuous blood pressure validated against Riva-Rocci/Korotkoff
Daan W Eeftinck Schattenkerk1, Johannes J van Lieshout, Anton H van den Meiracker
1Special Medical Care, Department of Internal Medicine, Academic Medical Center, University of Amsterdam, Amsterdam, The Netherlands.
Background:
The Finapres methodology offers continuous measurement of blood pressure (BP) in a noninvasive manner. The latest development using this methodology is the Nexfin monitor. The present study evaluated the accuracy of Nexfin noninvasive arterial pressure (NAP) compared with auscultatory BP measurements (Riva-Rocci/Korotkoff, RRK).
Methods:
In supine subjects NAP was compared to RRK, performed by two observers using an electronic stethoscope with double earpieces. Per subject, three NAP-RRK differences were determined for systolic and diastolic BP, and bias and precision of differences were expressed as median (25th, 75th percentiles). Within-subject precision was defined as the (25th, 75th percentiles) after removing the average individual difference.
Results:
A total of 312 data sets of NAP and RRK for systolic and diastolic BP from 104 subjects (aged 18-95 years, 54 males) were compared. RRK systolic BP was 129 (115, 150), and diastolic BP was 80 (72, 89), NAP-RRK differences were 5.4 (-1.7, 11.0) mm Hg and -2.5 (-7.6, 2.3) mm Hg for systolic and diastolic BP, respectively; within-subject precisions were (-2.2, 2.3) and (-1.6, 1.5) mm Hg, respectively.
Conclusion:
Nexfin provides accurate measurement of BP with good within-subject precision when compared to RRK.
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement