Abstract:
Medial opening wedge high tibial osteotmy (HTO) is often used to treat varus gonarthrosis in young, active, highly demanding patients, although it has many pitfalls, which were evaluated in a consecutive cohort of patients. A retrospective analysis of a consecutive series of 45 patients with 49 medial opening HTO for varus gonarthrosis using a spacer plate (Puddu I, Arthrex, USA) were included. A Chi square test was used to study the effect between the wedge size and complications. Complications occurred in 22 knees (45%). There was no significant difference between groups for individual complications; however, when combined, there were significantly more complications in the >10 mm wedge group (Chi square p = 0.05). The overall complication rate in this series was 45%. The majority were related to intrinsic instability at the osteotomy site (24%) and surgical technique (20%). The evaluated spacer provided inadequate stability.