Home sphygmomanometers: validation versus accuracy.
Tekin Akpolat1, Melda Dilek, Turkan Aydogdu
1Department of Nephrology, Ondokuz Mayis University School of Medicine, Samsun, Turkey. tekinakpolat@yahoo.com
Validated home blood pressure monitors are significantly more accurate than non-validated ones. This study highlights the importance of device validation for reliable hypertension management and accurate home blood pressure readings.
Area of Science:
- Cardiovascular Medicine
- Biomedical Engineering
- Public Health
Background:
- Accuracy and validation of home sphygmomanometers are critical for both clinical practice and the consumer market.
- A significant number of home blood pressure devices may not meet accuracy standards, potentially leading to misdiagnosis or inadequate treatment.
Purpose of the Study:
- To determine the prevalence of validated home sphygmomanometers available to consumers.
- To investigate the correlation between the validation status and the accuracy of these devices.
Main Methods:
- 896 home sphygmomanometers were collected from individuals at a hypertension clinic.
- 870 devices were suitable for analysis; validation status was checked against the dabl Educational Trust website.
- Accuracy was assessed in 554 automated sphygmomanometers, comparing validated and non-validated devices.
Main Results:
- Only 30% (22 of 74) of the evaluated devices were validated.
- A high rate of inaccuracy was found, with 72% (400 of 554) of automated sphygmomanometers being inaccurate.
- Validated devices demonstrated a significantly higher accuracy rate (68%) compared to non-validated devices (15%) (P<0.01).
Conclusions:
- Validated home sphygmomanometers are more likely to be accurate than non-validated devices.
- This study provides the first evidence linking the validation status to the accuracy of home blood pressure monitors.
- Improving patient education and market supervision, potentially with non-profit organization support, can reduce device-related errors in blood pressure monitoring.
More Related Videos
09:18Doppler Ultrasound-Based Leg Blood Flow Assessment During Single-Leg Knee-Extensor Exercise in an Uncontrolled Setting
Published on: December 15, 2023
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure