An In-House-Built and Light-Emitting-Diode-Based Photodynamic Therapy Device for Enhancing Verteporfin Cytotoxicity in a 2D Cell Culture Model
Published on: January 13, 2023
Evaluation of actual vs expected photodynamic therapy spot size
Tushar M Ranchod1, Alexander J Brucker, Chengcheng Liu
1Scheie Eye Institute, Department of Ophthalmology, University of Pennsylvania, Philadelphia, Pennsylvania 19104, USA.
Purpose:
To determine the accuracy of the photodynamic therapy (PDT) laser spot size on the retina as generated by 2 Food and Drug Administration (FDA)-approved lasers.
Design:
Prospective observational case series.
Methods:
Fundus photographs were taken of 1 eye of each of 10 subjects with the WinStation 4000 fundus photography system (OIS; Ophthalmic Imaging Systems, Sacramento, California, USA); disc size was calculated using OIS software. Slit-lamp photographs were taken of the PDT laser spot focused on the retina adjacent to the optic disc, using various spot sizes in combination with 3 different contact lenses and 2 different lasers. Spot size at the retina was determined by measuring the ratio of disc diameter to spot diameter in Adobe Photoshop (San Jose, California, USA) and applying this ratio to the OIS disc measurements.
Results:
Spot size at the retina averaged 87% of expected spot size for the Coherent Opal laser (Coherent Inc, Santa Clara, California, USA) and 104% of expected spot size for the Zeiss Visulas laser (Carl Zeiss Meditec Inc, Dublin, California, USA)(P = .002). Multivariate analysis demonstrated that percentage of expected spot size decreased with larger spot diameter (P = .01 for Coherent laser; P = .02 for Zeiss laser).
Conclusions:
PDT spot size at the retina appears to be consistently smaller than expected for the Coherent laser while the spot size was consistently within 10% of expected size for the Zeiss laser. The deviation from expected size increased with larger spot size using the Coherent laser.