Usefulness of a breath-holding acquisition method in PET/CT for pulmonary lesions
Toshiaki Yamaguchi1, Osamu Ueda, Hideyuki Hara
1Radiology Division, Hyogo Cancer Center, Akashi, Hyogo, 673-8558, Japan. tyamaguchi-rtec@umin.ac.jp
Objective:
To evaluate the usefulness of a breath-holding (BH) (18)F-2-fluoro-2-deoxy-D-glucose positron emission tomography ((18)F-FDG-PET) technique for PET/computed tomography (CT) scanning of pulmonary lesions near the diaphragm, where image quality is influenced by respiratory motion.
Methods:
In a basic study, simulated breath-holding PET (sBH-PET) data were acquired by repeating image acquisition eight times with fixation of a phantom at 15 s/bed. Free-breathing PET (FB-PET) was simulated by acquiring data even as moving the phantom at 120 s/bed (sFB-PET). Images with total acquisition times of 15 s, 30 s, 45 s, 60 s, and 120 s were generated for sBHPET. Receiver-operating characteristic (ROC) analyses and determination of the statistical significance of differences between sFB-PET images and sBH-PET images were performed. A total of 22 pulmonary lesions in 21 patients (12 men and 9 women, mean age 61.3 +/- 10.6 years, 10 benign lesions in 9 patients and 12 malignant lesions in 12 patients) were examined by FB-PET and BH-PET). For evaluation of these two acquisition methods, displacement of the lesion between CT and PET was considered to be a translation, and the statistical significance of differences in maximum standardized uptake value (SUV(max)) of the lesion was assessed using the paired t test.
Results:
In the basic study, sBH-PET images with acquisition times of 45 s, 60 s, and 120 s had significantly higher diagnostic accuracy than 120-s sFB-PET images (P < 0.05). In clinical cases, translation of the BH-PET images was significantly lower than that of the FB-PET images (benign: 5.29 +/- 4.02 mm vs. 11.79 +/- 8.27 mm, P = 0.005; malignant: 4.29 +/- 3.36 mm vs. 18.26 +/- 12.31 mm, P = 0.003). The SUV(max) of the lesions in the BH-PET images was significantly higher than that in the FB-PET images (benign: 2.40 +/- 0.86 vs. 2.20 +/- 0.85, P = 0.005; malignant: 4.84 +/- 2.16 vs. 3.75 +/- 2.11, P = 0.001).
Conclusions:
BH-PET provides images with better diagnostic accuracy, avoids image degradation owing to respiratory motion, and yields more accurate attenuation correction. This method is very useful for overcoming the problem of respiratory motion.
More Related Videos
02:09Multi-modal Pulmonary Imaging: Using Complementary Information from CT and Hyperpolarized 129Xe MRI to Evaluate Lung Structure-Function
Published on: April 12, 2024
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...