Airway humidification with a heated wire humidifier during high-frequency ventilation using Babylog 8000 plus in
Ken Nagaya1, Toshio Okamoto, Eiki Nakamura
1The Center for Maternity and Infant Care, Asahikawa Medical College, Midorigaoka-higashi, Asahikawa, Hokkaido, Japan. nagaya5p@asahikawa-med.ac.jp
Objective:
Little data are available on airway humidity during high-frequency ventilation (HFV). Our purpose is to evaluate the airway humidification during HFV.
Methods:
We examined the airway humidification and temperature in a neonatal HFV system using Babylog 8000 plus. The absolute humidity (AH), relative humidity (RH), and temperature at different sites and under different HFV conditions were compared with those during conventional intermittent positive pressure ventilation (IPPV).
Results:
The mean AH and RH at the patient end of the respiratory circuit under 37 degrees C in the humidification chamber (HC) during HFV were less than 35 mg/L and 65%, respectively, while those during IPPV were 42.3 mg/L and 96.8%, respectively. The humidification at the outlet of the HC was similar results. Moreover, during HFV an increase in the bias-flow of ventilator led to a further decrease in the humidity at the patient end of respiratory circuit and the outlet of HC. It was necessary to set the temperature in the HC at >39 degrees C to maintain adequate humidity at the HC and the patient end of respiratory circuit during HFV. An increase in the incubator temperature led to an increase in the temperature at the patient end of the respiratory circuit. The temperature at the patient end of the respiratory circuit was about 39-40 degrees C when the incubator temperature was 35-37 degrees C.
Conclusions:
The airway humidification at the patient end of respiratory circuit and the outlet of HC in HFV were poorer than those in IPPV. However, the adequacy of humidification and safety in HFV remain to be demonstrated in clinical practice.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Suctioning the Nasopharyngeal Airway
Equipment Required
Mechanical Ventilation I: Indication and Settings
Tracheostomy Suctioning II: Procedure
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...