Initial radiation experience evaluating early tolerance and toxicities in patients undergoing accelerated partial breast irradiation using the Contura Multi-Lumen Balloon breast brachytherapy catheter
Related Experiment Video
Updated: Jun 25, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Sheree Brown1, Mark McLaughlin1, Keith Pope1
1Department of Radiation Oncology, WellStar Kennestone Hospital, Marietta, GA.
Purpose:
We reviewed our institution's experience treating patients with the Contura Multi-Lumen Balloon (SenoRx, Inc., Irvine, CA) breast brachytherapy catheter to deliver accelerated partial breast irradiation.
Methods And Materials:
Forty-one patients treated with breast-conserving therapy received adjuvant radiation using the Contura catheter (34Gy in 3.4Gy fractions). Thirteen patients had Stage 0, 21 had Stage I, and 7 had Stage II breast cancer. Median followup was 8 months (range, 1-17).
Results:
Median, minimum skin spacing was 10mm (range, 2-17). Median, maximum skin doses (% of prescribed dose [PD]) were 99.7 (range, 57.1-124.1). Eight patients were treated with a skin spacing
Conclusion:
Adjuvant accelerated partial breast irradiation using the Contura Multi-Lumen Balloon catheter exhibited similar toxicities to standard single lumen balloon brachytherapy with improvements in dosimetric capabilities (i.e., reduced skin and rib doses and improved PTV_EVAL coverage).