Related Experiment Video
Updated: Jun 25, 2026

Microsurgical Creation of Giant Bifurcation Aneurysms in Rabbits for the Evaluation of Endovascular Devices
Published on: September 8, 2023
Embolization of a wide-necked basilar bifurcation aneurysm by double-balloon remodeling using HyperForm compliant
Hiroshi Shima1, Motohiro Nomura, Naoki Muramatsu
1Department of Neurosurgery, Yokohama Sakae Kyosai Hospital, 132 Katsura-cho, Sakae-ku, Yokohama, 247-8581 Japan. insula@ns.m.kanazawa-u.ac.jp
Abstract:
We report a patient with a wide-necked basilar bifurcation aneurysm that was difficult to address by standard coil embolization because the left and right posterior cerebral artery (PCA) branched from the base of the aneurysm. To preserve both PCAs, the aneurysm was endovascularly embolized using a double-balloon remodeling technique and HyperForm compliant balloon catheters (MicroTherapeutics, Irvine, CA, USA). The aneurysmal neck was not tightly embolized to avoid coil protrusion into the parent arteries. The present technique is an important adjunct to the embolization of wide-necked basilar bifurcation aneurysms.