Related Experiment Video
Updated: Jun 25, 2026

Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
A new low-level gamma-ray spectrometry system for environmental radioactivity at the underground laboratory
M Köhler1, D Degering, M Laubenstein
1Verein für Kernverfahrenstechnik und Analytik Rossendorf e.V. (VKTA), P.O. Box 510119, D-01314 Dresden, Germany. matthias.koehler@vkta.de
Abstract:
A low-level gamma-ray spectrometry system, based on an HPGe-detector with 92% relative efficiency recently installed in the underground laboratory Felsenkeller at 110 m water equivalent (w.e.) depth, is described. The integral background count rate normalised to the Ge-crystal mass in the energy range from 40 keV until 2.7 MeV of 0.034 s(-1)kg(-1) has been achieved by careful material selection of the detector construction material, a graded shielding construction and effective radon suppression. The detector is highly suitable for the effective surveillance of water for human consumption with decision thresholds for (226)Ra and (228)Ra in the order of some mBq L(-1).