Coherent scatter computed tomography for structural and compositional stone analysis: a prospective comparison with
Geoffrey R Wignall1, Ian A Cunningham, John D Denstedt
1Department of Surgery, Division of Urology, Schulich School of Medicine and Dentistry, The University of Western Ontario , London, Ontario, Canada.
Coherent scatter computed tomography (CSCT) offers superior stone analysis compared to infrared spectroscopy (IRS), accurately identifying stone composition and core structures without sampling bias.
Area of Science:
- Medical imaging
- Materials science
- Analytical chemistry
Background:
- Infrared spectroscopy (IRS) is a standard but destructive stone analysis method.
- IRS analyzes powdered samples, leading to sampling bias and potential underestimation or omission of components.
- IRS cannot provide structural composition, particularly of the stone core.
Purpose of the Study:
- To compare the efficacy of Coherent Scatter Computed Tomography (CSCT) with Infrared Spectroscopy (IRS) for stone analysis.
- To evaluate CSCT's ability to provide detailed structural and compositional analysis, including stone core imaging.
- To assess CSCT's advantage over IRS in avoiding sampling bias.
Main Methods:
- Consecutive patients with stone disease were recruited.
- Collected stones or fragments underwent both CSCT and IRS analysis.
- Comparison focused on bulk composition and core material identification.
Main Results:
- CSCT and IRS agreed on the primary component in 84.8% of samples.
- CSCT identified additional components in 88.8% of stones IRS deemed uniform.
- CSCT detected stone cores in 78.8% of samples, while IRS missed them in 21.2%.
Conclusions:
- CSCT provides superior, quantitative stone analysis without sampling error, analyzing the entire specimen.
- CSCT offers excellent structural imaging, including detailed analysis of core composition.
- CSCT is a more comprehensive method for stone analysis than IRS.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
The ATR process begins by directing a beam...
Imaging Studies III: Computed Tomography
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Atomic Fluorescence Spectroscopy