Published on: December 15, 2014
Towards optimized acquisition scheme for multiprojection correlation imaging of breast cancer
Amarpreet S Chawla1, Robert S Saunders, Swatee Singh
1Department of Radiology, Duke Advanced Imaging Laboratories, Duke University, Durham, NC 27705, USA. asc14@duke.edu
Rationale And Objectives:
Correlation imaging (CI) is a form of multiprojection imaging in which multiple images of a patient are acquired from slightly different angles. Information from these images is combined to make the final diagnosis. A critical factor affecting the performance of CI is its data acquisition scheme, because nonoptimized acquisition may distort pathologic indicators. The authors describe a computer-aided detection (CADe) methodology to optimize the acquisition scheme of CI for superior diagnostic accuracy.
Materials And Methods:
Images from 106 subjects were used. For each subject, 25 angular projections of a single breast were acquired. Projection images were supplemented with a simulated 3-mm three-dimensional lesion. Each projection was then processed using a traditional CADe algorithm at high sensitivity, followed by the reduction of false-positives by combining the geometric correlation information available from the multiple images. The performance of the CI system was determined in terms of free-response receiver-operating characteristic curves and the areas under receiver-operating characteristic curves. For optimization, the components of acquisition, such as the number of projections and their angular span, were systematically changed to investigate which of the many possible combinations maximized the obtainable CADe sensitivity and specificity.
Results:
The performance of the CI system was improved by increasing the angular span. Increasing the number of angular projections beyond a certain number did not improve performance. Maximum performance was obtained between 7 and 10 projections spanning a maximum angular arc of 45 degrees .
Conclusion:
The findings suggest the existence of an optimum acquisition scheme for CI of the breast. CADe results confirmed earlier predictions on the basis of observer models. An optimized CI system may be an important diagnostic tool for improved breast cancer detection.
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging