Precise Lung Density Quantification with a Physics-based CT Harmonizer
Saman Sotoudeh-Paima1,2, David A Lynch3, Stephen M Humphries3
1Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Laboratories, Department of Radiology, Duke University School of Medicine, 2424 Erwin Rd, Hock Plaza, Ste 302, Durham, NC 27705.
Abstract:
Purpose To develop a physics-based image harmonization method that transforms images into a reference quality index of noise, spatial resolution, and lung volume and to evaluate its performance for improving reproducibility of lung density measurement. Materials and Methods This retrospective analysis of Genetic Epidemiology of Chronic Obstructive Pulmonary Disease (COPDGene) study data included participants who underwent chest CT with full-dose (200 mAs) and reduced-dose (40-80 mAs) protocols during the same visit between November 2014 and July 2017. A harmonization algorithm was developed to reduce variations in lung density by adjusting spatial resolution, noise, and lung volume in sequence, aligning scans to a reference condition based on full-dose, soft-kernel reconstruction. The percentage of lung voxels less than -950 HU (LAA-950) and the 15th percentile of the lung density histogram (Perc15) were calculated from scans. Improvements in bias, limits of agreement, and reproducibility coefficients were evaluated, and harmonizer performance was compared with that of two existing techniques: volume-adjusted lung density (VALD) and median filtering followed by VALD (MF-VALD). Results A total of 1159 participants (mean age ± SD, 65 years ± 9; 586 male) were studied. Across all imaging conditions, the harmonization technique improved the Perc15 reproducibility coefficient 4.8-fold, from 35.6 HU ± 0.7 to 7.4 HU ± 0.2. The harmonizer outperformed VALD and MF-VALD across the full-dose and reduced-dose scans, with reproducibility coefficients improving from 32.4 HU ± 0.7 before harmonization to 30.1 HU ± 0.3 with VALD, 10.3 HU ± 0.3 with MF-VALD, and 7.7 HU ± 0.2 with the harmonizer. Conclusion This physics-based technique harmonized CT images to a reference quality index, improving the reproducibility of lung density metrics. Keywords: CT, Thorax, Lung, Physics, Chronic Obstructive Pulmonary Disease Clinical trial registration no. NCT00608764 Supplemental material is available for this article. © RSNA, 2026.
More Related Videos
07:53Author Spotlight: Advancing 3D Modeling for Enhanced Diagnosis and Treatment of Pulmonary Nodules in Early-Stage Lung Cancer
Published on: October 13, 2023
03:38Unilateral Lung Volume Analysis Using Micro-CT for Enhanced Assessment of Pulmonary Fibrosis in Preclinical Models
Published on: June 20, 2025
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT