Reproducibility of regional pulse wave velocity in healthy subjects
1Department of Medical Equipment Information, Vision College of Jeonju, Jeonju, Korea.
Background/Aims:
Despite the clinical importance and widespread use of pulse wave velocity (PWV), there are no standards for pulse sensors or for system requirements to ensure accurate pulse wave measurement. We assessed the reproducibility of PWV values using a newly developed PWV measurement system.
Methods:
The system used in this study was the PP-1000, which simultaneously provides regional PWV values from arteries at four different sites (carotid, femoral, radial, and dorsalis pedis). Seventeen healthy male subjects without any cardiovascular disease participated in this study. Two observers performed two consecutive measurements in the same subject in random order. To evaluate the reproducibility of the system, two sets of analyses (within-observer and between-observer) were performed.
Results:
The means+/-SD of PWV for the aorta, arm, and leg were 7.0+/-1.48, 8.43+/-1.14, and 8.09+/-0.98 m/s as measured by observer A and 6.76+/-1.00, 7.97+/-0.80, and 7.97+/-0.72 m/s by observer B, respectively. Betweenobserver differences for the aorta, arm, and leg were 0.14+/-0.62, 0.18+/-0.84, and 0.07+/-0.86 m/s, respectively, and the correlation coefficients were high, especially for aortic PWV (r=0.93). All the measurements showed significant correlation coefficients, ranging from 0.94 to 0.99.
Conclusions:
The PWV measurement system used in this study provides accurate analysis results with high reproducibility. It is necessary to provide an accurate algorithm for the detection of additional features such as flow wave, reflection wave, and dicrotic notch from a pulse waveform.
Related Concept Videos
Pulse
The pulse serves as a clinical indicator...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring pulse
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement