Comparing noninvasive blood pressure monitoring technologies: the challenge of between-arm differences
Simon T Vistisen1, Alexandre Joosten2, Johannes Enevoldsen1
1Department of Clinical Medicine, Aarhus University, Aarhus, Denmark; Department of Anesthesiology and Intensive Care, Aarhus University Hospital, Aarhus, Denmark.
Abstract:
Continuous noninvasive blood pressure monitoring technologies are increasingly being evaluated against invasive arterial pressure measurements. Berkholz and colleagues demonstrate that a finger-cuff system satisfies International Organization for Standardization (ISO) performance criteria for mean arterial and diastolic blood pressure, although not for systolic blood pressure, and it appeared to show closer agreement with invasive monitoring than oscillometric measurements. However, because oscillometric measurements were necessarily obtained from the contralateral arm, physiological inter-arm blood pressure differences may have contributed to the observed differences in agreement with the invasive reference standard. In this editorial, we describe how the study supports mean arterial pressure monitoring with a finger-cuff system and why it does not establish superiority over conventional oscillometric monitoring.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure