Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Objective vs subjective vault measurement after myopic implantable collamer lens implantation
José F Alfonso1, Carlos Lisa, Ana Palacios
1Fernández-Vega Ophthalmological Institute, University of Oviedo, Oviedo, Spain. j.alfonso@fernandez-vega.com
Purpose:
To evaluate the relationship between subjective measurements of vault and objective values measured with Visante optical coherence tomography (OCT) in eyes receiving an implantable contact lens (ICL) for myopia correction.
Design:
Observational cross-sectional study.
Setting:
Fernández-Vega Ophthalmological Institute, Oviedo, Spain.
Patients:
Four hundred and fifty-two eyes from 246 patients were elected to be implanted with a Visian ICL V4 (STAAR Surgical Inc, Monrovia, California, USA).
Observation Procedures:
Subjective and objective measurements of vault after implantation of ICL.
Main Outcome Measures:
Subjective vault classified in 5 levels assessed using an optical section during slit-lamp examination. Objective vault was measured with Visante OCT (Carl Zeiss Meditec Inc, Dublin, California, USA).
Results:
Average values of objective and subjective vault were 414 +/- 228 microm and 2.1 +/- 1.0, respectively and both parameters were highly correlated (r = 0.82; P < .001). Differences in average objective vault were statistically significant among the 5 groups of subjective vault (P < .001). Subjective vault 0 corresponded to a mean OCT value of 62 +/- 49 microm with 99% confidence interval (CI) [38; 86] microm; eyes with vault 1 to 203 +/- 93 microm with 99% CI [176; 230] microm; eyes with vault 2 to 402 +/- 131 microm with 99% CI [378; 425] microm; eyes with vault 3 to 594 +/- 146 microm, 99% CI [554; 633] microm; and vault 4 to 794 +/- 182 microm with 99% CI [713; 875] microm.
Conclusions:
Subjective and objective values of vault are highly correlated. In 99% of cases within the CI, objective values for eyes subjectively classified within a certain level vary within a narrow interval (+/-25 to 80 microm) around the mean value, and this interval is characteristic of each subjective level.