A bioabsorbable everolimus-eluting coronary stent system (ABSORB): 2-year outcomes and results from multiple imaging
Patrick W Serruys1, John A Ormiston, Yoshinobu Onuma
1Thorax Center, Erasmus Medical Center, Rotterdam, Netherlands.
Background:
Drug-eluting metallic coronary stents predispose to late stent thrombosis, prevent late lumen vessel enlargement, hinder surgical revascularisation, and impair imaging with multislice CT. We assessed the safety of the bioabsorbable everolimus-eluting stent (BVS).
Methods:
30 patients with a single de-novo coronary artery lesion were followed up for 2 years clinically and with multiple imaging methods: multislice CT, angiography, intravascular ultrasound, derived morphology parameters (virtual histology, palpography, and echogenicity), and optical coherence tomography (OCT).
Findings:
Clinical data were obtained from 29 of 30 patients. At 2 years, the device was safe with no cardiac deaths, ischaemia-driven target lesion revascularisations, or stent thromboses recorded, and only one myocardial infarction (non-Q wave). 18-month multislice CT (assessed in 25 patients) showed a mean diameter stenosis of 19% (SD 9). At 2-year angiography, the in-stent late loss of 0.48 mm (SD 0.28) and the diameter stenosis of 27% (11) did not differ from the findings at 6 months. The luminal area enlargement on OCT and intravascular ultrasound between 6 months and 2 years was due to a decrease in plaque size without change in vessel size. At 2 years, 34.5% of strut locations presented no discernible features by OCT, confirming decreases in echogenicity and in radiofrequency backscattering; the remaining apparent struts were fully apposed. Additionally, vasomotion occurred at the stented site and adjacent coronary artery in response to vasoactive agents.
Interpretation:
At 2 years after implantation the stent was bioabsorbed, had vasomotion restored and restenosis prevented, and was clinically safe, suggesting freedom from late thrombosis. Late luminal enlargement due to plaque reduction without vessel remodelling needs confirmation.
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion, evaluates...