Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon
G S Stergiou1, P Lourida, D Tzamouranis
1Hypertension Center, Third University Department of Medicine, Sotiria Hospital, Athens, Greece. gstergi@med.uoa.gr
Abstract:
Oscillometric devices are being widely used for ambulatory, home and office blood pressure (BP) measurement. However, even successfully validated oscillometric devices fail to provide accurate measurements in some patients. This study investigated the prevalence, the reproducibility and the characteristics of the phenomenon of unreliable oscillometric BP (UOBP) measurement. A total of 5070 BP measurements were obtained simultaneously (Y connector) using a professional oscillometric device (BpTRU) and a mercury sphygmomanometer in 755 patients (1706 visits). UOBP readings were defined as those with >10 mm Hg difference (systolic or diastolic) between the two methods. UOBP was found in 15% of systolic and 6.4% of diastolic BP measurements. In all, 18% of the participants had UOBP in their first but not their second visit, or the reverse. However, 49% of these participants had at least one more UOBP visit after their second visit within the study database. Patients with persistent UOBP were more likely to be female and had lower arm circumference. The systolic BP discrepancy between the two methods was associated with pulse pressure (r=0.41) and inversely with diastolic BP (r=0.40) and arm circumference (r=0.30), whereas the diastolic discrepancy with diastolic BP (r=0.61) and inversely with pulse pressure (r=0.32). There was a consistent significant trend for larger systolic BP discrepancy and smaller diastolic from the lower to the higher pulse pressure quintile (P<0.0001). A decreasing arm circumference was a significant predictor of persistent UOBP. These data suggest that the UOBP measurement is particularly common, not very reproducible and mainly affected by pulse pressure and arm circumference.
Related Concept Videos
Measurement of Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...