Comparison of dynamic contour tonometry and Goldmann applanation tonometry and their relationship to corneal
Dina H Erickson1, Denise Goodwin, Michael Rollins
1Pacific University College of Optometry, Forest Grove, Oregon 97116, USA. derickson@pacificu.edu
Background:
Accurate intraocular pressure (IOP) measurement is essential in diagnosing and managing glaucoma. Dynamic contour tonometry (DCT) is less dependent on corneal properties, such as thickness, elasticity, and rigidity, than Goldmann applanation tonometry (GAT). This study examined the relationship between GAT and DCT as well as their relationship with corneal properties and ocular pulse amplitude (OPA).
Methods:
GAT, DCT, OPA, pachymetry, refractive error, and corneal curvature measurements were obtained on 115 healthy volunteers.
Results:
Participants with thicker corneas (>or=580 microm) had higher IOP measurements with GAT than DCT (P = 0.005). Those with thinner corneas (
Conclusion:
DCT provides IOP measurements that are less dependent on corneal factors than GAT, aiding in diagnosis and treatment of patients with ocular hypertension and glaucoma. Additional studies are necessary to examine the relationship between OPA, refractive error, and IOP and its possible association with increased incidence of glaucoma in myopic patients.