Related Experiment Video
Updated: Jun 24, 2026

Thermal Measurement Techniques in Analytical Microfluidic Devices
Published on: June 3, 2015
Application of the thermal flash technique for low thermal diffusivity micro/nanofibers
Michael T Demko1, Zhenting Dai, Han Yan
1Department of Mechanical and Aerospace Engineering, Case Western Reserve University, Cleveland, Ohio 44106, USA.
Abstract:
The thermal flash method was developed to characterize the thermal diffusivity of micro/nanofibers without concern for thermal contact resistance, which is commonly a barrier to accurate thermal measurement of these materials. Within a scanning electron microscope, a micromanipulator supplies instantaneous heating to the micro/nanofiber, and the resulting transient thermal response is detected at a microfabricated silicon sensor. These data are used to determine thermal diffusivity. Glass fibers of diameter 15 microm had a measured diffusivity of 1.21x10(-7) m(2)/s; polyimide fibers of diameters 570 and 271 nm exhibited diffusivities of 5.97x10(-8) and 6.28x10(-8) m(2)/s, respectively, which compare favorably with bulk values.