Effect of mattress deflection on CPR quality assessment for older children and adolescents
Akira Nishisaki1, Jon Nysaether, Robert Sutton
1Department of Anesthesiology and Critical Care Medicine, The Children's Hospital of Philadelphia (CHOP), USA. NISHISAKI@email.chop.edu
Insights
Chest compression depth during CPR is often overestimated on hospital beds due to mattress compression. This significantly reduces the proportion of adequate compressions, potentially impacting CPR outcomes in children.
Area of Science:
- Pediatric Critical Care Medicine
- Cardiopulmonary Resuscitation (CPR) Quality Assessment
- Biomedical Engineering
Background:
- Appropriate chest compression (CC) depth is crucial for effective cardiopulmonary resuscitation (CPR) and improved patient outcomes.
- Hospital environments, particularly compliant mattresses, can interfere with accurate CC depth assessment during resuscitation efforts.
- The impact of mattress compression on CPR quality metrics in pediatric patients remains incompletely understood.
Purpose of the Study:
- To quantify the effect of mattress compression on the assessment of CPR quality in children.
- To determine the overestimation of CC depth caused by mattress deflection during CPR.
- To evaluate the impact of mattress compression on the proportion of CCs meeting established guidelines.
Main Methods:
- A force and deflection sensor (FDS) was utilized during actual CPR events in a pediatric intensive care unit and emergency department.
- CPR events were reconstructed using a manikin with an identical mattress/backboard/patient configuration to measure mattress deflection.
- CC depth was measured using the FDS placed between the rescuer's hands and the patient's chest, with simultaneous accelerometer data to calculate mattress deflection.
Main Results:
- Twelve CPR events (14,487 CCs) from 11 pediatric patients were analyzed, with compressions performed on ICU beds and stretchers.
- Mattress compression led to an overestimation of CC depth by 13+/-4mm on ICU beds and 4+/-1mm on stretcher beds.
- After adjusting for mattress compression, the proportion of adequate CCs meeting CPR guidelines decreased significantly from 88.4% to 31.8% on ICU beds (p<0.001) and 86.3% to 64.7% on stretchers (p<0.001).
Conclusions:
- Chest compressions performed on non-rigid surfaces like hospital mattresses may not achieve adequate depth.
- Force and deflection sensors can overestimate CC depth by up to 28% on ICU beds and 10% on stretcher beds due to mattress compression.
- The findings highlight the critical need to account for mattress properties when assessing CPR quality in pediatric patients to ensure effective resuscitation.
Unlabelled:
Appropriate chest compression (CC) depth is associated with improved CPR outcome. CCs provided in hospital are often conducted on a compliant mattress. The objective was to quantify the effect of mattress compression on the assessment of CPR quality in children.
Methods:
A force and deflection sensor (FDS) was used during CPR in the Pediatric Intensive Care Unit and Emergency Department of a children's hospital. The sensor was interposed between the chest of the patient and hands of the rescuer and measured CC depth. Following CPR event, each event was reconstructed with a manikin and an identical mattress/backboard/patient configuration. CCs were performed using FDS on the sternum and a reference accelerometer attached to the spine of the manikin, providing a means to calculate the mattress deflection.
Results:
Twelve CPR events with 14,487 CC (11 patients, median age 14.9 years) were recorded and reconstructed: 9 on ICU beds (9296 CC), 3 on stretchers (5191 CC). Measured mean CC depth during CPR was 47+/-8mm on ICU beds, and 45+/-7 mm on stretcher beds with overestimation of 13+/-4mm and 4+/-1mm, respectively, due to mattress compression. After adjusting for this, the proportion of CC that met the CPR guidelines decreased from 88.4 to 31.8% on ICU beds (p<0.001), and 86.3 to 64.7% on stretcher (p<0.001). The proportion of appropriate depth CC was significantly smaller on ICU beds (p<0.001).
Conclusion:
CC conducted on a non-rigid surface may not be deep enough. FDS may overestimate CC depth by 28% on ICU beds, and 10% on stretcher beds.
Related Concept Videos
Cardiopulmonary Resuscitation I: Adult
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.