Related Experiment Video
Updated: Jun 24, 2026

Quantification of the Immunosuppressant Tacrolimus on Dried Blood Spots Using LC-MS/MS
Published on: November 8, 2015
Analysis of cyclosporin A in dried blood spots using liquid chromatography tandem mass spectrometry
A J Wilhelm1, J C G den Burger, R M Vos
1VU University Medical Center, Laboratory of Pharmacology and Pharmacy, HV, Amsterdam, Netherlands.
Abstract:
Dried blood spot sampling is a promising and patient friendly alternative for venous sampling. A liquid chromatography tandem mass spectrometer assay was developed for analyzing cyclosporin A in dried blood spots. Linearity ranged from 25 to 1440 microg/L. Within and between run accuracy and precision were within limits. The developed assay has a negligible matrix-effect and a recovery of 97%. The dried blood spots were stable during a period of at least 17 days in the refrigerator. The developed assay is suitable for analyzing cyclosporin A in dried blood spots.