Evaluation of a point-of-care glucose meter for general use in complex tertiary care facilities
Pak Cheung R Chan1, Mary Rozmanc, Isolde Seiden-Long
1Department of Clinical Pathology, Sunnybrook Health Sciences Centre, Toronto, Ontario, Canada M4N 3M5. pc.chan@sunnybrook.ca
Objective:
To evaluate the performance of the StatStrip glucose meter from Nova Biomedical for use in complex tertiary care facilities.
Methods:
Performance evaluation was conducted in 6 clinical locations involving nurse end-users. Imprecision (CV)<5% was considered acceptable. Ten substances were examined for potential analytical interferences at two levels and 3 glucose concentrations. Inaccuracy was determined by 386 paired glucose differences between glucose meter and laboratory reference analyses. Acceptability was based on CSLI/ISO15197 criteria. Potential clinical impact was classified under a 5-level severity scheme.
Results:
Imprecision varied from CV 2.4 to 4.9%. Significant interference was observed only for free hemoglobin at 10 g/L. 97% of paired glucose meter-laboratory differences met CSLI/ISO15197 criteria. All discrepant results had low severity scores.
Conclusion:
The StatStrip glucose meter demonstrated acceptable imprecision and inaccuracy, and was relatively free from common interferences. It should be a good candidate for general use in complex tertiary care facilities.
Related Concept Videos
Type II Diabetes Mellitus III: Clinical Manifestations and Diagnosis
Hyperglycemia
Diabetes: Symptoms, Diagnosis, and Complications
Automated Microbial Diagnostics
Amperometry: Overview
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...