Assessing the performance of the Whisperflow continuous positive airway pressure generator: a bench study
1Intensive Care Unit, Bradford Teaching Hospitals NHS Foundation Trust, Duckworth Lane, Bradford BD9 6RJ, UK.
Background:
There are few data describing the performance of the Whisperflow continuous positive airway pressure (CPAP) generator.
Methods:
(i) (a) A static test of 11 Whisperflow devices examining maximum flow generation with no load and with 2.5, 5, 7.5, 10, 15, and 20 cm H2O valve loading, at varying Fi(O2). (b) CPAP valves (Accu-peep(R), Vital Signs, Totowa, NJ, USA) were tested by measuring mean upstream pressure at varying flows in five valves (2.5, 5, 7.5, 10, 15, and 20 cm H2O). (ii) We measured the mean and minimum inspiratory mask pressure generated by a representative Whisperflow device in a model of spontaneous respiration. Measurements were made with combinations of Fi(O2), ventilatory frequency, tidal volume, and valve loading similar to those encountered in clinical practice.
Results:
(i) (a) The flow generated by the Whisperflow valves decreases with increasing valve load and increasing Fi(O2) (from 140 to 20 litre min(-1)). (b) The CPAP valves maintain the required pressure within acceptable limits against varying flow. (ii) At all permutations, the mean inspiratory mask pressure was significantly lower than that required. At high inspiratory flow rates, the minimum inspiratory pressure approached atmospheric pressure.
Conclusions:
The Whisperflow may not perform as expected. Clinicians should be cautious when using this device, particularly with high Fi(O2) and CPAP valve load. The flow setting should be set at maximum. Failure of CPAP therapy may be due to failure of the generator. Further in vivo data are required.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...