Increases in endotracheal tube resistance are unpredictable relative to duration of intubation
Alison M Wilson1, Dana M Gray2, John G Thomas2
1Departments of Surgery, West Virginia University, Morgantown, WV.
Background:
Accumulated secretions after intubation can affect the resistance of an endotracheal tube (ETT). Our objective was to measure extubated patient tubes and size-matched controls to evaluate differences in resistance.
Methods:
New ETTs, with internal diameters of 7.0 through 8.5 mm, were tested as controls to establish the resistance of each size group as measured by pressure drop. Measurements were obtained using a mass flowmeter and pressure transducer. Pressure drop was measured at three flow rates. Seventy-one patient ETTs were evaluated after extubation by an identical method and compared with controls.
Results:
In each control group, pressure drop was tightly clustered with low variation and no overlap between sizes. A total of 73 to 79% of the patient ETTs had a pressure drop of > 3 SDs of size-matched controls at all flow rates. Pressure drop in 48 to 56% (across three flow rates) of extubated tubes was equivalent to the next smaller size of controls. At 60 and 90 L/min, 10% and 15% of patient tubes, respectively, had the pressure drop of a control tube three sizes smaller. The pressure drop was unpredictable relative to the duration of intubation.
Conclusions:
Organized secretions can significantly increase resistance as measured by the pressure drop of ETTs. The degree of change was highly variable, occurs in all sizes, and was unrelated to the duration of intubation. The performance of an ETT may be comparable to new tubes one to four sizes smaller. This may impact the tolerance of ventilator weaning.
Related Concept Videos
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs during...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...